Pressure Specification
Gauge ranges (referenced to atmosphere) | |
Pressure range | Accuracy | Total uncertainty 10o to 30oC (50o to 86oF) for 1 year | Overpressure protection2 |
bar | psi | %FS | %FS | %FS |
-1 to 1 | -14.5 to 15 | 0.0185 | 0.025 | 150 |
-1 to 2 | -14.5 to 30 | 0.0185 | 0.025 | 150 |
-1 to 7 | -14.5 to 100 | 0.0185 | 0.025 | 150 |
-1 to 10 | -14.5 to 150 | 0.0185 | 0.025 | 150 |
-1 to 20 | -14.5 to 300 | 0.0185 | 0.025 | 150 |
Pressure media
ºÎ½Ä ¼º ¾ø´Â ´ëºÎºÐ °¡½º¿Í ȣȯÀÌ °¡´É.
Pressure connection
µµ±¸ ¾øÀÌ ºü¸¥ ¸ÂÃã ¿¬°á. G1/8Female°ú 1/8NPT Female ¾î´ðÅÍ¿Í ÇÔ²² Á¦°ø
External pressure modules
IDOS (Áö´ÉÇü µðÁöÅÐ Ãâ·Â ¼¾¼) ¾Ð·Â ¸ðµâÀº ±â±âÀÇ ÃøÁ¤ ¹üÀ§¸¦ È®ÀåÇÏ´Â USB º¯È¯±â P / N IO620-USB-IDOS ÅëÇØ ¿¬°á°¡´É
Multiple parameter display capability
4 °³ÀÇ µ¿½Ã Àбâ â Ãִ븦 Ç¥½ÃÇϵµ·Ï ±¸¼º °¡´É : ¾Ð·Â °ª, Àü±â ÃøÁ¤, Àü±â ¼Ò½º, IDOS ¿ÜºÎ ¾Ð·Â ¸ðµâ
Electrical Specificaiton
Electrical measurement and source |
Measure mode | Total uncertainty 10oC to 30oC (50o to 86oF) for 1 year | Additional error -10oC to 10oC & 30oC to 50oC (-10oC to 10oF & 30oC to 50oF) |
Resolution |
| %Rdg + %FS | %Rdg/¡ÆC + %FS/¡ÆC | |
DC Voltage | | | | | |
+/- 200 mV | 0.015 | 0.01 | 0 | 0.0005 | 0.001 |
+/- 2000 mV | 0.015 | 0.01 | 0 | 0.0005 | 0.01 |
+/- 20 V | 0.015 | 0.01 | 0 | 0.0005 | 0.00001 |
+/- 30 V | 0.015 | 0.01 | 0 | 0.0005 | 0.0001 |
Current | | | | | |
+/- 20 mA | 0.015 | 0.01 | 0 | 0.0005 | 0.0001 |
+/- 55 mA | 0.018 | 0.01 | 0 | 0.0005 | 0.0001 |
Source mode | | | | | |
DC Voltage | | | | | |
10V (Fixed, 25mA max.) | 0 | 0.1 | 0 | 0 | 0.001 |
24V (Fixed, 25mA max.) | 0 | 1.0 | 0 | 0 | 0.001 |
Current | | | | | |
0 to 24 mA | 0.018 | 0.01 | 0 | 0.0005 | 0.001 |
0 to 24 mA (loop power) | 0.018 | 0.01 | 0 | 0.0005 | 0.001 |